Sign in →

Test Code D Dimer Qt D-DIMER QUANTITATIVE

Patient Preparation

No special patient preparation required

Special Collection Instructions

Properly filled 9:1 ratio (blood to anticoagulant); 3.2% NaCitrate tube

Collection Method

Venous Draw; properly filled 9:1 ratio (blood to anticoagulant)

Minimum Volume

0.5 of plasma from a properly filled 1.8 mL or 2.7 mL NaCitrate (light blue) tube

Container

Light Blue Sodium Citrate tube

Preferred Specimen

2.7 mL Lt Blue (3.2% NaCitrate) tube

Acceptable Specimen(s)

1.8 mL Lt Blue (3.2% NaCitrate) tube

Reference or Target Ranges

Normal Range:

0.00-0.50 mg/L FEU

Negative Cut-off Range:

0.50 mg/L FEU

Reportable Units

FEU (Fibrinogen Equivalent Units)

Critical Value

Not defined for this assay

Reasons for Rejection

Hemolysis Threshold:

100 mg/dL

Icterus Threshold:

15 mg/dL

Lipemia Threshold:

600 mg/dL

Other:

Short draws, over filling tubes, clotted, contaminated, Improperly labeled, samples >24 hours from collection

Specimen Stability

Ambient:

24 Hours

Frozen:

15 Days

Performance Information

Days and Time Performed

24/7

Expected Turn Around Time

Stat: 1 Hour, Routine: 4 Hours

Stat Availability

Yes

Performing Bench

CRMC Coag

Methodology/Method Description

CS-2500

Methodology/Method Description

CS-2500 Chromogenic assay

CPT Codes

85379